AEROSET
®
System Troubleshooting Guide
94816-107—November 2004
Verification Procedures/Adjustments/Alignments
Lamp Gain and A/D Check
Perform the following procedure to check lamp gains and A/D
reads. This will determine if the lamp should be replaced or if
additional troubleshooting is necessary.
Materials Required
None
Procedure
1. Ensure the system status is READY.
2. Remove one cuvette segment.
3. Select the
Maintenance Utilities
icon in the Action Area
of the Main Display. The
MAINTENANCE UTILITIES
screen displays.
4. Select the
Rxn Area
tab. The
Rxn Area
page displays.
Figure 16: MAINTENANCE UTILITIES Screen, Rxn Area Page
5. Using the Rxn Crsl buttons, rotate the Reaction Carousel
to place the empty position in the optics light path.
6. Select
Shutter <ON>
to open the shutter.
Rxn Crsl
Shutter
Data