Startup/Shutdown Setting Default Screens
Default views of the AEROSET software v1.02ER000 automatic
Startup/Shutdown Setting
screen are shown below. Also
included is a table that describes fields on the
Startup/Shutdown Setting
screen which can be affected if the
System Configuration Software Disk v3.0 is loaded after
upgrade to software v1.02ER000.
Figure 10: Startup/Shutdown (Startup-1) Settings Screen, Default
View
* None of the Automatic START UP Options are customer
defined. Each of these settings should match the defaults
listed in this table.
Automatic START UP Options
Option
Startup-1
Startup-2
Startup-3
Change Bath-water
X
X
X
Wash Cuvettes
X
X
X
Wash Probes by:
Water
Alk
Water
• Water
• Alk (Wash)
• Acid (Wash)
• Alk + Acid
Wash ISE by:
ICAL
EW + ICAL
• ICAL (IRef)
• EW (I-Clean)
• EW + ICAL
Change MCC
(Drain & Fill IRef Cup)
X
X
Probe Air Purge
(Flush Water Lines)
X
X
X