Hardware
AEROSET
®
System Troubleshooting Guide
227
94816-107—November 2004
Observed Problems—Hardware
Figure 14: Clot Detection Pressure Sensors
Erroneous Error Log Messages related to the Clot Detection system:
341 Invalid Status Was Received from PM Board
342 Sample Aspiration Error Occurred (A-Line)
343 Sample Aspiration Error Occurred (B-Line)
344 Unable to Remove Sample Probe Obstruction (A-Line)
345 Unable to Remove Sample Probe Obstruction (B-Line)
NOTE:
Refer to the following illustrations for details.
Probable Cause(s)
Corrective Action(s)
Sample probe is obstructed or
damaged.
Clean or replace sample probe.
Sample arm tubing is crimped.
Replace the sample arm tubing.
Pressure sensor tubing is loose or
crimped.
Reseat or replace the pressure sensor tubing.
Pressure sensor is defective.
Contact your Abbott Reprepresentative to
replace the pressure sensor.