PROPRIETARY INFORMATION
CONFIDENTIAL MATERIAL
Assays/Reagents
Not for Customer Distribution
56
AEROSET
®
System Troubleshooting Guide
94816-107—November 2004
THC50
Antibody to THC (R1) + G-6-PD-THC (R2) (Active)
G-6-PD-THC:Antibody complex (Inactive)
G-6-PD-THC:Antibody complex (Inactive) + UC
G-6-PD-THC (Active) + UC:Antibody complex
G-6-PD-THC (Active) + NAD + Glucose-6-phosphate
NADH + 6-phosphogluconate
THC =
∆
9
-Tetrahydrocannabinol
G-6-PD = Glucose-6-phosphate dehydrogenase from bacterial source
UC = Urine cannabinoids
NADH = Nicotinamide adenine dinucleotide
Use Multi DOA Cutoff
calibrator in duplicate
for calibration. Cutoff
calibrator value is
zero on
CALIBRATOR/
CONTROL
screen.
Certain types of
plastics adsorb
cannabinoids (THC)
from solutions,
decreasing the
concentration. To
prevent adsorbtion:
• Fill AEROSET
sample cups with
600 µL to
minimize
surface-to-volume
ratio.
• Do not use
disposable
polyethylene
transfer pipettes to
dispense urine
specimens,
calibrator, or
controls.
• Do not redispense
solutions from a
pipette into a
calibrator,
controls, or
specimen
container
(includes
redispensing after
wetting pipette
tip).
• Immerse pipette
tip below surface
of the liquid only
far enough to
aspirate the
needed volume.
Assay
Name
Reaction
Tips
* MULTIGENT
®
assay
Содержание AEROSET
Страница 19: ...AEROSET System Troubleshooting Guide 19 94816 107 November 2004 Assays Reagents Assays Reagents ...
Страница 221: ...AEROSET System Troubleshooting Guide 221 94816 107 November 2004 Section 10 Hardware Hardware ...
Страница 245: ...AEROSET System Troubleshooting Guide 245 94816 107 November 2004 Flowcharts Troubleshooting Flowcharts ...