PROPRIETARY INFORMATION
CONFIDENTIAL MATERIAL
Not for Customer Distribution
Assays/Reagents
AEROSET
®
System Troubleshooting Guide
31
94816-107—November 2004
Example of Absorbance Mode Results Printout
Each assay value in this example would be compared to its
cutoff calibrator absorbance to determine if the result is
positive or negative.
The following are examples only and do not demonstrate
actual absorbance values.
The Barbiturates and Cannabinoids (THC50) assays in this
example would be reported as positive.
Amphetamine/Methamphetamine, Benzodiazepines, Cocaine,
Methadone, Opiates 300, Opiates 2,000, Phencyclidine, and
Propoxyphene would be reported as negative.
10/17/01
13:01
SN A7602220
SID 007
Name
John Doe
POS
5 ( 2)
PID
Date/Time Run
10/17/01 13:01
ERR
NAME
L/H
RESULT
ERR
NAME
L/H
RESULT
PCP
0.3237
AmpM
0.2776
Prpx
0.3870
Coc
0.3212
Opi3
0.2341
Benz
0.2702
Meth
0.3988
Barb
0.2926
Assay
Sample
Absorbance
Cutoff
Absorbance
Result
Amphetamine/
Methamphetamine
0.2776
0.2778
Negative
Barbiturates
0.2926
0.2923
Positive
Benzodiazepines
0.2702
0.2705
Negative
Cannabinoids 50
0.3206
0.3206
Positive
Cocaine
0.3212
0.3232
Negative
Methadone
0.3988
0.3989
Negative
Opiates 300
0.2341
0.2378
Negative
Opiates 2,000
0.2781
0.2789
Negative
Phencyclidine
0.3237
0.3286
Negative
Propoxyphene
0.3870
0.3889
Negative