7. OPERATION VIA BRAIN TERMINAL
IM 1E7B0-02E
7-11
7.2.11 Automatic Two Range Switching (Refer to 6.3.9)
Display
Description
PULSE OUT
ALARM OUT
BI DIRECTION
AUTO 2 RANGES
LOW ALARM
TOTAL SWTICH
Pulse output
Alarm output
Forward / reverse flow measurement
Automatic two range switching
Alarm output at low flow limits
Totalization switch
PARAM
B09:HIGH MF
1.0000
B10:OUTPUT FUNC
AUTO 2 RANGES
B11:4-20 ALM OUT
2.4mA OR LESS
DATA
ESC
DIAG
DATA
ESC
DIAG
PARAM
B31 :BI DIREC HYS
B33 :FOR. SPAN2
B34 :AUTO RNG HYS
Select "AUTO 2 RANGES" (Automatic two range transfer) in parameter
number "B10".
Forward second range can be set by calling up parameter No. B33 .
Setting restrictions: First range
≤
2nd range.
Further second range span should be set in the same number of places of decimals
as first range span in parameter No. B03 .
Example: First range : 1.000 then second range should be
4.000
Hysteresis width at switching range can be set in parameter No. B34 .
It is the rate (%) of first range span.
Content
Code
7.2.12 Alarm Output at Low Flow Limits (Flow Switch) (Refer to 6.3.10)
Display
Description
PULSE OUT
ALARM OUT
BI DIRECTION
AUTO 2 RANGES
LOW ALARM
TOTAL SWTICH
Pulse output
Alarm output
Direct / reverse flow measurement
Automatic two range transfer
Alarm output at low flow limits
Totalization switch
PARAM
B09:HIGH MF
1.0000
B10:OUTPUT FUNC
LOW ALARM
B11:4-20 ALM OUT
2.4mA OR LESS
DATA
ESC
DIAG
DATA
ESC
DIAG
PARAM
B34 :AUTO RNG HYS
B36 :LOW ALARM
B37 :L. ALARM HYS
Select "LOW ALARM " (Alarm output at low flow limits) in parameter
number "B10 ".
Set the comparison level in parameter number "B36 ".
Set the hysteresis in parameter "B37 ".