IM 1E7D0-01E
3-16
3. INSTALLATION
3.4.4
Connnection to AE14 Magnetic
Flow Converter
Flow Tube Model AE100D, AE200D and AE300D
Connect the flow tube and converter in the following
method.
EX2
EX1
A
B
C
SA
(See Note below)
A
SB
(See Note below)
C
EX1
EX2
B
A
C
EX1
EX2
B
Dedicated Cable
AM011-4
Converter
Terminals
Connected Terminals
Flowtube AE100D,
AE200D and AE300D
Note: Terminate those shielding wire
terminals using insulating tape.
Excitation Cable
Magnetic Flow Tube
AE100D, AE200D and AE300D
Magnetic Flow Converter
AE14
P
- P
+ I- I+
L
/+ N
/-
P
L
S
/A
L
M
O
U
T
C
U
R
O
U
T
P
O
W
E
R
S
U
P
P
LY
EX1 EX2 C SA
A
B SB
!
Figure 3.4.4
Connection
3.4.5
Wiring Ports
Please select the most suitable standard of wiring
procedure for the wiring ports by customer’s own.
A : Using the Waterproof Gland
IMPORTANT
To prevent water or condensate from entering
the converter housing, waterproof glands are
recommended. Do not over-tighten the glands or
damage to the cables may result. Tightness of
the gland can be checked by confirming that the
cable is held firmly in place.
Gasket
Tightening
gland
G1/2 female
Gasket
Tightening gland
G1/2 female
Waterproof gland
Optional code : /ECG
Waterproof gland with union joint
Optional code : /ECU
F030406.EPS
Figure 3.4.5
Waterproof Gland
B : Insulation Check
After wiring is completed, check insulation of the
following terminals together with the wiring under the
condition of the converter side wiring terminals
disconnected.
Between terminal EX1 and terminal A, B, C
Between terminal C and terminal A, B
Between terminal A and terminal B
Between terminal EX1 and ground
Between terminal EX2 and ground
All insulation measurements must be performed with a
500V megger. Insulation resistances must be 100M
Ω
or more each.
IMPORTANT
Be sure to disconnect the cables at the terminal
of the converter when checking insulation.
C : Conduit Wiring
In case of conduit wiring, please use the waterproof
gland to prevent water flowing through the conduit
pipe into the wiring connection.
Please slope the conduit pipe down, and install a drain
valve at the low end of the vertical pipe.
Please open the drain valve regularly.
Drain valve
F030407.EPS
Figure 3.4.6
Conduit Wiring