![YOKOGAWA AE100D Instruction Manual Download Page 201](http://html.mh-extra.com/html/yokogawa/ae100d/ae100d_instruction-manual_911083201.webp)
IM 1F6A0-01E
5-14
5. BASIC OPERATING PROCEDURES
5.6.2 Display
The HART Communicator automatically searches for
YEWFLO on the 4 to 20 mA loop when it is turned on.
When the HART Communicator is connected to the
YEWFLO, it displays “Online” menu as shown below.
(If YEWFLO is not found, the communicator displays the
message “No Device Found. Press OK....” Press the OK ‘F4’
function key and the main menu appears. Retry after
confirming the connection with the YEWFLO.)
Manufacturer’s Field Device
Type Code
Tag (8 Characters)
<1>
<2>
<3>
<4>
<5>
Function keys
appears when the
voltage level of the
battery is low
The arrow mark
corresponding to the
pressed key appears.
The highlighting cursor
DYF:
Online
1 Device setup
2 PV % rnge 0.0875kg/min
3 PV AO 5,600mA
4 PV URV 0.8746kg/min
HELP
SAVE
HOME
ENTER
F1
F2
F3
F4
F050601.EPS
Figure 5.6.1 Display
<1>
appears and flashes during communication between
the HART Communicator and the YEWFLO. At Burst
mode*,
appears.
<2> The current display menu title appears.
<3> Each item in menu of <2> appears.
<4>
and/or
appear when the items are scrolled out of
the display.
<5> On any given menu, the label appearing above a
function key indicates the function of that key for the
current menu.