6. FUNCTION AND DATA SETTINGS
IM 1E7B0-02E
6-8
6.3.5
Limiting Current Output During Alarm Occurrence
• The current output and display values during alarming can be selected in advance.
Display
Description
Set a value for current output to be used during alarms in parameter number "11"
Code
Contents
00
01
02
03
2.4mA or less
4.0mA
Hold
21.6mA or more
Default : 00
(2.4 mA or less)
6.3.6 Reversing Flow Direction
• The flow direction is set to "FORWARD" at the factory. This function enables to set
flow direction from "FORWARD" to "REVERSE".
Display
Description
The flow direction can be set in parameter No."14".
Code
Contents
00
01
Forward direction
Reverse direction to flow arrow
Default : 00
6.3.7
Limiting on Current Output
(1) 4 to 20 mA Low Cut Output (Current Output near by 0% Range)
• This function makes it possible to reduce fluctuations in the 0% region to reduce it to
0%.
Display
Description
Setting range : 0 to 10%
0.5% 0.5%
4mA
Output
Set value
Input
Hysteresis : 1% fixed
Default : 0(%)
5.6mA(10%)