7. OPERATION VIA BRAIN TERMINAL
IM 1E7B0-02E
7-9
7.2.6
Damping Time Constant (Refer to 6.3.4)
Display
Description
PARAM
B01:TAG NO
B02:DAMPING
3.0 s
B03:FLOW SPAN
1.00 m/s
DATA
ESC
DIAG
Set the value in parameter number "B02."
Setting Range : 0.1 to 200.0 seconds.
7.2.7
Current Output during Alarm Occurrence (Refer to 6.3.5)
Display
Description
PARAM
B10:OUTPUT FUNC
PULSE OUT
B11:4-20 ALM OUT
2.4mA OR LESS
B12:POWER FREQ
50.03Hz
DATA
ESC
DIAG
Set the value for current output to be used during alarms in parameter
"B11."
(Default : 2.4 mA or less)
2.4mA OR LESS
4.0mA
HOLD
21.6mA OR MORE
2.4 mA or less
4.0 mA
Fixed to output when alarm is
generated.
21.6 mA or more
Content
Code
7.2.8
Reversing Flow Direction (Refer to 6.3.6)
Display
Description
PARAM
B13:VELOCITY CHK
5 m/s
B14:FLOW DIR
REVERSE
B60:SELF CHECK
GOOD m/s
DATA
ESC
DIAG
The flow direction can be set in parameter number "B14"
Code
Description
FORWARD
REVERSE
Forward direction
Reverse direction to flow arrow
7.2.9
Limiting Current Output (Refer to 6.3.7)
(1) 4 to 20 mA low cut output (Current Output near by 0% Range)
Display
Description
PARAM
G01:4-20 LOW CUT
10 %
G02:4-20 LOW LMT
-20 %
G03:4-20 H LMT
110 %
DATA
ESC
DIAG
Setting range : 0 to 10%
0.5% 0.5%
4mA
Output
Set value
Hysteresis:1% fixed
5.6mA(10%)
Default:0%