![YOKOGAWA AE100D Instruction Manual Download Page 315](http://html.mh-extra.com/html/yokogawa/ae100d/ae100d_instruction-manual_911083315.webp)
IM 01C21B01-01E
10-1
10. GENERAL SPECIFICATIONS
10. GENERAL SPECIFICATIONS
10.1
Standard Specifications
Refer to IM 01C22T02-01E for Fieldbus communi-
cation type marked with “
”.
Performance Specifications
See General Specifications sheet, GS 01C21B01-
00E, GS 01C21B03-00E, and GS 01C21B04-00E.
Functional Specifications
Span & Range Limits
EJA110A and EJA130A:
L
M
Measurement
Span
and Range
Span
Range
Span
Range
inH
2
O
(/D1)
2 to 40
–40 to 40
4 to 400
–400 to
400
mbar
(/D3)
5 to 100
mmH
2
O
(/D4)
50 to 1000
–100 to 100
10 to 1000
–1000 to
1000
–1000 to
1000
100 to 10000
–10000 to
10000
kPa
0.5 to 10
–10 to 10
1 to 100
–100 to
100
H
Span
Range
20 to 2000
–2000 to
2000
50 to 5000
–5000 to
5000
0.05 to
5 kgf/cm
2
–5 to
5 kgf/cm
2
5 to 500
–500 to
500
T1001.EPS
V*
0.14 to
14 MPa
20 to
2000 psi
1.4 to
140 bar
1.4 to
140
kgf/cm
2
–0.5 to
14 MPa
–71 to
2000 psi
–5 to
140 bar
–5 to
140 kgf/cm
2
Span
Range
* For Wetted parts material code other than S, the ranges
are 0 to 14 MPa, 0 to 2000 psi, 0 to 140 bar, and 0 to
140 kgf/cm
2
.
EJA120A:
E
Measurement
Span
and Range
Span
Range
inH
2
O
(/D1)
0.4 to 4
–4 to 4
mbar
(/D3)
1 to 10
mmH
2
O
(/D4)
10 to 100
–10 to 10
–100 to 100
kPa
0.1 to 1
–1 to 1
T1002.EPS
URL is define as the Upper Range Limit from the table
above.
Zero Adjustment Limits:
Zero can be fully elevated or suppressed, within
the Lower and Upper Range Limits of the capsule.
External Zero Adjustment “
”:
External zero is continuously adjustable with
0.01% incremental resolution of span. Span may
be adjusted locally using the digital indicator with
range switch.
Output “
”:
Two wire 4 to 20 mA DC output with digital
communications, linear or square root program-
mable. BRAIN or HART FSK protocol are superim-
posed on the 4 to 20 mA signal.
Failure Alarm:
Output status at CPU failure and hardware error;
Up-scale: 110%, 21.6 mA DC or more(standard)
Down-scale: -5%, 3.2 mA DC
Note: Applicable for Output signal code D and E
Damping Time Constant (1st order):
The sum of the amplifier and capsule damping
time constant must be used for the overall time
constant. Amp damping time constant is adjustable
from 0.2 to 64 seconds.
Time Constant (approx. sec)
0.4
0.3
0.3
0.3
0.2
L
M
H
V
E
T1003.EPS
Capsule (Silicone Oil)
Ambient Temperature Limits:
* Safety approval codes may affect limits.
EJA110A, EJA130A
–40 to 85
°
C (–40 to 185
°
F)
–30 to 80
°
C (–22 to 176
°
F) with LCD Display
EJA120A
–25 to 80
°
C (–13 to 176
°
F)
Process Temperature Limits:
* Safety approval codes may affect limits.
EJA110A, EJA130A
–40 to 120
°
C (–40 to 248
°
F)
EJA120A
–25 to 80
°
C (–13 to 176
°
F)
Maximum Overpressure:
See General Specifications sheet.
Working Pressure Limits (Silicone Oil)
Maximum Pressure Limit:
See ‘Model and Suffix Codes’
Minimum Pressure Limit:
Capsule
Pressure
L, M, H (EJA110A)
See Figure 1.
M, H (EJA130A)
See Figure 1.
E (EJA120A)
–50 kPa {–7.25 psig}