![YOKOGAWA AE100D Instruction Manual Download Page 320](http://html.mh-extra.com/html/yokogawa/ae100d/ae100d_instruction-manual_911083320.webp)
IM 01C21B01-01E
10-6
10. GENERAL SPECIFICATIONS
10.3
Optional Specifications
Item
Description
Code
Factory Mutual (FM)
CSA Explosionproof Approval
Explosionproof for Class I, Division 1, Groups B, C and D
Dustignitionproof for Class II/III, Division 1, Groups E, F and G
Division2 ‘SEALS NOT REQUIRED’ , Temp. Class : T4, T5, T6 Encl Type 4x
Max. Process Temp.: T4; 120
⬚
C (248
⬚
F), T5; 100
⬚
C (212
⬚
F), T6; 85
⬚
C (185
⬚
F)
Amb. Temp.:–40 to 80
⬚
C (–40 to 176
⬚
F)
CF1
Canadian Standards
Association (CSA)
CSA Intrinsically safe Approval
Class I, Groups A, B, C and D Class II and III, Groups E, F and G
Encl Type 4x, Temp. Class: T4, Amb. Temp.: –40 to 60
⬚
C (–40 to 140
⬚
F)
Vmax=30 V, Imax=165 mA, Pmax=0.9 W, Ci=22.5 nF, Li=730
H
CS1
CENELEC ATEX (KEMA) Intrinsically safe Approval
II 1G EEx ia IIC T4, Amb. Temp.: –40 to 60
⬚
C(–40 to 140
⬚
F)
Ui=30 V, Ii=165 mA, Pi=0.9 W, Ci=22.5 nF, Li=730
H
KS2
FM Intrinsically safe Approval
Intrinsically Safe for Class I, Division 1, Groups A, B, C & D, Class II, Division 1,
Groups E, F & G and Class III, Division 1 Hazardous Locations.
Nonincendive for Class I, Division 2, Groups A, B, C & D, Class II, Division. 2,
Groups E, F & G, and Class III, Division 1 Hazardous Locations.
Enclosure: “NEMA 4X”, Temp. Class: T4, Amb. Temp.: –40 to 60
⬚
C (–40 to 140
⬚
F)
Intrinsically Safe Apparatus Parameters
[Groups A, B, C, D, E, F and G]
Vmax=30 V, Imax=165 mA, Pmax=0.9 W, Ci=22.5 nF, Li=730
H
[Groups C, D, E, F and G]
Vmax=30 V, Imax=225 mA, Pmax=0.9 W, Ci=22.5 nF, Li=730
H
FS1
CENELEC ATEX (KEMA) Flameproof Approval
II 2G EExd IIC T4, T5, T6
Amb. Temp.: T5; –40 to 80
⬚
C ( –40 to 176
⬚
F), T4 and T6; –40 to 75
⬚
C( –40 to 176
⬚
F)
Max. process Temp.: T4; 120
⬚
C (248
⬚
F), T5; 100
⬚
C (212
⬚
F), T6; 85
⬚
C (185
⬚
F)
KF2
FM Explosionproof Approval
Explosionproof for Class I, Division 1, Groups B, C and D
Dust-ignitionproof for Class II/III, Division 1, Groups E, F and G
Hazardous (classified) locations, indoors and outdoors ( NEMA 4X )
Temperature class: T6
Amb. Temp.:–40 to 60
⬚
C (–40 to 140
⬚
F)
FF1
Standards Association of
Australia (SAA)
SU1
SAA Flameproof, Intrinsically safe and Non-sparking Approval
Ex d IIC T4/T5/T6, IP67 class I, Zone 1, Amb. Temp. : –40 to 80
⬚
C (–40 to 176
⬚
F)
Max. Process Temp.: T4; 120
⬚
C (248
⬚
F), T5; 100
⬚
C (212
⬚
F), T6; 85
⬚
C (185
⬚
F)
Ex ia IIC T4, IP67 class I, Zone 0
Ex n IIC T4, IP67 class I, Zone 2
Ui=30 V DC, Ii=165 mA DC, Wi=0.9 W, Amb. Temp.: –40 to 60
⬚
C (–40 to 140
⬚
F)
FU1
Combined FF1and FS1
Combined KF2, KS2, Type n, and Dust Approval
Type n: II 3G Ex nL IIC T4, Amb. Temp.: –40 to 60
°
C (–40 to 140
°
F)
Ui=30 V, Ci=22.5 nF, Li=730
µ
H
Dust: II 1D Maximum surface temperature T65
°
C (Tamb.: 40
°
C), T85
°
C (Tamb.: 60
°
C),
and T105
°
C (Tamb.: 80
°
C)
KU2
CENELEC ATEX
CU1
Combined CF1 and CS1
T1008.EPS