![YOKOGAWA AE100D Instruction Manual Download Page 39](http://html.mh-extra.com/html/yokogawa/ae100d/ae100d_instruction-manual_911083039.webp)
IM 1E7D0-01E
6-3
6. HAZARDOUS DUTY TYPE INSTRUMENT
(4) Maintenance and Repair
WARNING
The instrument modification or parts replacement
by other than authorized representative of
YOKOGAWA Electric Corporation or
YOKOGAWA Corporation of AMERICA is
prohibited and will void Canadian Standards
Explosionproof Certification.
Conduit
Magnetic Flow Converter
Sealing Fitting
HAZARDOUS LOCATIONS
50 cm Max.
F0602.EPS
Figure 6.3.1 Conduit Wiring
6.4 SAA
WARNING
Only trained persons use this instrument in
industrial locations .
(1) Technical Data
• SAA :
SA Certificate No.
AUS Ex 3210X
Type of Protection
Ex dm(ia)II C T6 IP67
Electrode Circuit Um
250V ac/dc
Excitation Circuit130Vp 72/75Hz
Temperature Class
T6
T5
T4
T3
Process Temperature
70
85
120
150
°
C
(2) Installation
WARNING
• All wiring shall comply with local installation
requirements and local electrical code.
• In hazardous locations, the cable entry devices
shall be of a certified flameproof type, suitable
for the conditions of use and correctly installed.
(3) Operation
WARNING
• Open circuit before opening the covers.
• Take care not to generate mechanical spark
when access to the instrument and peripheral
devices in hazardous locations.
(4) Maintenance and Repair
WARNING
The instrument modification or parts replacement
by other than authorized representative of
Yokogawa Electric Corporation is prohibited and
will void the certification.