6. FUNCTION AND DATA SETTINGS
IM 1E7B0-02E
6-7
6.3.3
Resetting for Totalization Display
• This function is used to reset or preset totalization values of the display.
• Hold down the SHIFT key for more than 2 seconds while the totalization values of
the flow rate are displayed to set the totalization value to the value set in parameter
number "E5".
Display
Description
Select totalization enable in parameter number "E4"
Code
Description
00
01
Totalization preset inhibit
Totalization preset enable
Set the totalization preset value in parameter number "E5". The initial value
is 0, if it is no setting, the function is zero setting.
Hold down the SHIFT key for more than 2 seconds while the totalization
values of the flow rate reading are displayed to set the totalization value to
the value set in parameter number "E5"
Default : 00 (inhibit)
Default : 0
SET SHIFT INC
6.3.4
Damping Time Constant
• The time constant can be changed by setting the parameter No."02" to suppress a
fluctuation or change a response time.
• The time constant influences to flow rate, pulse output and internal totalization.
However, in case "00" is selected in parameter number "n1", the pulse output and
internal totalization are not influenced by it.
Display
Description
Set the value in parameter number "02".
Setting Range : 0.1 to 200.0 seconds.
Default : 3 seconds