7. OPERATION VIA BRAIN TERMINAL
IM 1E7B0-02E
7-6
7.2 Setting Parameters
The three parameter must be set to obtain a correct signal.
Nominal size, flow span and meter factor must be set.
In these three factors, Nominal size(unit: mm) and meter facror are set before
shipment.
7.2.1
Setting Flow Span
Example: Flow span 10.0 m
3
/h
PARAM
B01:TAG NO
B02:DAMPING
3.0 S
B03:FLOW SPAN
1.00 m/s
SET
B03:FLOW SPAN
1.00 m/s
0
CLR
ESC
DEL
MENU
A:DISPLAY
B:SET
C:ADJUST
D:DISP SEL
E:TOTAL SET
F:PULSE SET
HOME
ADJ
ESC
SET
ESC
DIAG
DATA
Call up the menu screen and move
the inverse video bar to “B : SET.”
Press the ENTER key to call the
screen displaying the parameter sub
items. Move the inverse video bar
to “B03 : FLOW SPAN”.
Pressing the ENTER key causes the
data setup screen to be displayed. If
the security screen is displayed, enter
the security code.*
SET
B03:FLOW SPAN
10.0 m/s
CLR
ESC
DEL
OK
NO
SET
B02:DAMPING
3.0 S
B03:FLOW SPAN
10.0 m/s
B04:FLOW UNIT
m
ESC
DIAG
DATA
The screen displaying the parameter
sub-items is displayed again. Feed
the page using the arrow key and
move the inverse video bar to the
item “B04 : FLOW UNIT.”
Pressing the ENTER key causes the
data display to blink. After
confirming it, press the ENTER key
once more and the above screen
appears. If the data are correct,
press the F4 key.
Enter “10.0” using the alphanumeric
keys. (Enter 10.0 only for the sub
item “B03 : FLOW SPAN”.) The
unit, m
3
/h is selected at items B04
and B05.
SET
B04:FLOW UNIT
m
3
OK
NO
SET
B04:FLOW UNIT
m
< km
3
>
<
m
3
>
< l >
< cm
3
>
ESC
SET
B03:FLOW SPAN
10.0 m/s
B04:FLOW UNIT
m
3
B05:TIME UNIT
/s
ESC
DIAG
DATA
Pressing the ENTER key causes the
data setup screen to be displayed.
Move the inverse video bar to “m
3
.”
If the security screen appears, enter
the security code* to display the
above screen.
* km
3
:10
3
⫻
m
3
cm
3
:(10
-2
⫻
m)
3
Press the ENTER key twice in
succession and the above screen
appears. Check that the data are
correct, and press the F4 key.
The screen displaying the parameter
sub-items appears again. Feed the
page using the arrow key and move
the inverse video bar to “B05 :
TIME UNIT”.
Setting is completed. The screen
returns to the preceding screen when
the F4 (ESC) key is pressed.
Press the ENTER key twice in
succession and the above screen
appears. Check that the data are
correct, and press the F4 key.
Pressing the ENTER key causes
the data setup screen to be
displayed. Move the inverse video
bar to “/h”.
SET
B04:TIME UNIT
/h
OK
NO
PARAM
B03:FLOW SPAN
10.0 m
3
/h
B04:FLOW UNIT
m
3
B05:TIME UNIT
/h
ESC
SET
CLR
HOME
SET
B05:TIME UNIT
/s
< /s >
< /d >
<
/h
>
< /m >
ESC
SET
B03:FLOW SPAN
1.00 m/s
10.0
*For entry of the security code, see IM 1C0A11-01E.
Inverse video bar
NOTE