![YOKOGAWA AE100D Instruction Manual Download Page 36](http://html.mh-extra.com/html/yokogawa/ae100d/ae100d_instruction-manual_911083036.webp)
IM 1E7D0-01E
5-12
5. OUTLINE
Sanitary, Remote Magnetic Flow Tube AE200DH
Unit: mm
(approx. inch)
F28.EPS
ø86(3.4)
48
(1.9)
øD1
L2
L2
1.7
(0.07)
1.7
(0.07)
H2
H1
H
36
(1.4)
Ground
Terminal
(M4)
109
(4.3)
øD
L
øD2
Type
Sanitary(H)
PFA lining
Size code
Size
Lining
Face-to-face
Ferrule length
Ferrule outer dia.
Ferrule inner dia.
Clamp outer dia.
Height
Height
Height
Weight kg(lb)
202
25(1)
176(6.9)
42.8(1.7)
25.4(1.0)
23(0.9)
74(2.9)
212(8.3)
126(5.0)
49(1.7)
3.5(7.7)
186(7.3)
41.3(1.6)
38.1(1.5)
35.7(1.4)
90(3.5)
235(9.3)
149(5.8)
61(2.4)
4.6(10.1)
206(8.1)
44.3(1.7)
50.8(2.0)
47.8(1.9)
103(4.1)
248(9.8)
162(6.4)
67(2.6)
6.0(13.2)
246(9.7)
44.3(1.7)
76.3(3.0)
72.3(2.8)
131(5.2)
277(10.9)
191(7.5)
81(3.2)
9.9(21.8)
204
40(1.5)
205
50(2)
208
80(3)
L
L2
D1
D2
D
H
H1
H2
272(10.7)
44.3(1.7)
101.6(4.0)
97.6(3.8)
171(6.7)
329(13.0)
243(9.6)
111(4.4)
15.4(33.8)
210
100(4)
■
SIZING DATA
15
25
40
50
80 100
150
200 250 300 350
Size: mm
400
Flow velocity
(m/s)
0.5
1.0
5.0
10.0
0.3
Flow velocity
(ft/s)
Flowrate (GPM)
33
5.0
1.0
10
2.0
1
SI Units
10
5
1
1,000
100
10
0.1
0.01
0.005
10,000
1.0
1,000
100
10
Size (inch)
0.1
0.01
0.5
2
1.5
1
3
4
6
12
10
8
1416
0.4
0.2
0.1
English Units
Flowrate (m3/h)
Size: inch
2.5
F29.EPS
* Measurable flow velocity is from 0 m/s.
■
GASKET
Please use compressed non-asbestos fiber gasket, PTFE gasket or the gasket which has
equal elasticity. In case of optional code/FRG, please use rubber gasket or others which has