6. FUNCTION AND DATA SETTINGS
IM 1E7B0-02E
6-4
6.2 Power Frequency
In case of DC power supply version, setting power frequency is required.
Display
Description
Set the value in area where the frequency differs in
parameter number "12".
Setting Range : 47.00 to 53.00 Hz.
Default : 50.00Hz
6.3 Other Functions and Settings
6.3.1
Pulse Output
(1) Pulse Output Overview
• By setting a pulse weight, a scaled pulse is transmitted to external counters and
measuring instruments.
Pulse Output Overview
Item
Content
Output
specifications
Connecting
terminals
Pulse width
Output rate
Transistor contact output (contact capacity is 30 V DC, 200 mA)
P+, P-
When using these for pulse output, alarm output or status output
are not available as the terminals are used commonly.
Selection: DUTY 50%, 0.5, 1, 20, 33, 50, 100 ms
Min. 0.0001 P/s Max. 1000 P/s
*
See page "4.4.6 Connecting external instruments" for information on how to connect external instruments.
P+, P- terminals are for common use with pulse, alarm and other status output
functions. Therefore, in case this function is used, other functions are not available
to use.
(2) Procedures for Setting Pulse Output
Example setting :
10 liter
output per pulse in a flow rate span of
m
3
/h
Display
Description
Select "Pulse output" in parameter number "10".
Code
Content
00
Pulse output
01
02
03
04
05
Alarm output
Forward/reverse flow measurement
Automatic two range switching
Alarm output at low flow limit
Totalization switch
IMPORTANT
NOTE