7. OPERATION VIA BRAIN TERMINAL
IM 1E7B0-02E
7-7
7.2.2
Power Frequency
In case of DC power supply version, setting power frequency is required.
Display
Description
Set value in areas where the frequency differs in "B12 : POWER FREQ".
Default:50.00Hz.
Setting Range : 47.00 to 53.00Hz
PARAM
B10:OUTPUT FUNC
PULSE OUT
B11:4-20ALM OUT
2.4mA OR LESS
B12:POWER FREQ
50.03Hz
DATA
ESC
DIAG
7.2.3
Pulse Output (Refer to 6.3.1)
Example setting : 10 L output per pulse in a flow rate span of
m
3
/h
PARAM
F01:PULSE UNIT
m UNIT/P
F02:PULSE SCALE
10 m UNIT/P
F03:PULSE LOWCUT
3. %
Display
Description
Code
Volume unit
n UNIT/P
m UNIT/P
m UNIT/P
UNIT/P
k UNIT/P
M UNIT/P
P ULSE/s
Volume unit in that for the flow rate span
×
10
-9
Volume unit in that for the flow rate span
×
10
-6
Volume unit in that for the flow rate span
×
10
-3
Volume unit in that for the flow rate span
×
1
Volume unit in that for the flow rate span
×
10
3
Volume unit in that for the flow rate span
×
10
6
Number of pulses output per second at 100% of
output
Code
Pulse width
50%DUTY
0.5ms
1ms
20ms
33ms
50ms
100ms
Example) When pulses are to be output per same liter with the flow rate span
of m
3
/h, select "m UNIT/P" since a L(liter) = 10
-3
×
m
3
Set the low cut range percentage in parameter "F03".
Range of setting : (0 to 100% of span)
The pulse output terminals (P+, P-) are also used for alarms. Select "Pulse
output" in parameter number "B10".
Select the volume unit for the pulse weight in parameter number "B04".
Select the pulse width in parameter number "F04"
Set the pulse weight "10 ( L )" in parameter number "F02".
*Since parameter number "F02" is a 5-digit data item, scrolling is
necessary to display all the data. Mind the decimal point when seting are
made. (The decimal point can be moved if required.)
(Max. of 1000P/s
Min. 0.0001P/s)
(Max. of 1000P/s
Min. 0.0001P/s)
(Max. of 500P/s
Min. 0.0001P/s)
(Max. of 25P/s
Min. 0.0001P/s)
(Max. of 15P/s
Min. 0.0001P/s)
(Max. of 10P/s
Min. 0.0001P/s)
(Max. of 5P/s
Min. 0.0001P/s)
PARAM
B10:OUTPUT FUNC
PULSE OUT
B11:4-20ALM OUT
2.4mA OR LESS
B12:POWER FREQ
50.03Hz
DATA
ESC
DIAG
DATA
ESC
DIAG
PALAM
F02:PULSE UNIT
10 m UNIT/P
F03:PULSE LOWCUT
3. %
F04:PULSE WIDTH
50% DUTY
DATA
ESC
DIAG
ESC
PULSE OUT
ALARM OUT
BI DIRECTION
AUTO 2 RANGES
LOW ALARM
TOTAL SWTICH
Pulse output
Alarm output
Forward / reverse flow measurement
Automatic two range switching
Alarm output at low flow limits
Totalization switch
Code
Content
PARAM
N01:TOTAL/PULSE
DAMP
N02:OUTPUT MODE
ON
N03:RATE LIMIT
5. %
DATA
ESC
DIAG
Normally, these are all required settings.
The following settings are made depending on the application that is used.
Select instantaneous flow rate or flow rate after damping data for the
pulse output.
(The damping time value is the value set in "02".)
Default : DAMP
Set parameter "02" to "OFF" when the pulse output transistor is to be
off-active.
IMPORTANT
* The "N" item can be opened by entering
"55" in parameter number "L02".