7. OPERATION VIA BRAIN TERMINAL
IM 1E7B0-02E
7-8
7.2.4
Display of Internal Totalization Values (Refer to 6.3.2)
Example : 10 L output per pulse in a flow rate span of
m
3
/h
Display
Description
Code
Volume unit
n UNIT/P
UNIT/P
m UNIT/P
UNIT/P
k UNIT/P
M UNIT/P
PULSE/s
Volume unit used in that for the flow rate span
×
10
-9
Volume unit used in that for the flow rate span
×
10
-6
Volume unit used in that for the flow rate span
×
10
-3
Volume unit used in that for the flow rate span
×
1
Volume unit used in that for the flow rate span
×
10
3
Volume unit used in that for the flow rate span
×
10
6
Number of pulse output per second at 100% of output
Example) When pulses are to be output per same liter with the flow rate
span of
m
3
/h, select “m UNIT/P” since a L(liter) =10
-3
×
m
3
Set the pulse weight “10 (L)” in parameter number “E02.”
Set the low cut range percentage in parameter “E03.” Range of
setting : (0 to 100% of span)
Select instantaneous flow rate or flow rate after damping for the pulse
output.
(The damping time value is the value set in “B02.” )
Default : DAMP (damped flow rate data)
Select the volume unit for the pulse weight in parameter number “E01.”
PARAM
E01:TOTAL UNIT
m UNIT/P
E02:TOTAL SCALE
10 m UNIT/P
E03:TOTAL LOWCUT
3. %
DATA
ESC
DIAG
PARAM
N01:TOTAL/PULSE
DAMP
N02:OUTPUT MODE
ON
N03:RATE LIMIT
5. %
DATA
ESC
DIAG
* The "N" item can be opened by entering "55" in parameter number "L02".
7.2.5
Resetting for Totalization Display (Refer to 6.3.3)
Display
Description
Code
Description
ENABLE
INHIBIT
Totalization presetting enabled
Totalization presetting disabled
Select totalization enable in parameter number “E04.”
Set the totalization preset value in parameter number “E05.”
The default is 0 and if the setting is omitted, the parameter functions
reset to 0.
If the ENT key is pressed twice while “A30 : TOTAL” is displayed on
the setting screen, the totalization value (A30 : TOTAL) will be
replaced with the values set in “E05.”
PARAM
E04:TOTAL SET
ENABLE
E05:TL SET VALUE
0
E10:TOTAL SWITCH
0
DATA
ESC
DIAG
PARAM
A10:FLOW RATE(%)
50 .0 %
A20:FLOW RATE
50 .0 m
3
/h
A30:TOTAL
12345
DATA
PRNT
ESC
DIAG
SETTING
A30:TOTAL
0
CLR
ESC
ENTER