6. FUNCTION AND DATA SETTINGS
IM 1E7B0-02E
6-12
6.3.10 Alarm Output at Low Flow Limit (Flow Switch)
•
In case flow rate decrease under set level, an status signal is output.
P+, P– terminals are for common use with pulse, alarm and other status output
functions. Therefore, in case this function is used, other functions are not
available to use.
Display
Description
Default : 00
(Default : 5%)
(Default : -10%)
Alarm output at low flow limits can be selected “04” in
parameter No. “10”.
The Low Limit value can be set in parameter No. “36”
as parcentage for 4 to 20 mA.
Hysteresis width is set in parameter No. “37”.
Code
00
01
02
03
04
05
Pulse output
Alarm output
Forward / reverse flow measurement
Automatic two range switching
Alarm output at low flow limit
Totalization switch
Content
Output
Hysteresis
Status
Output
On
low flow
limits
value.
When using reversed status (ON/OFF)
is required, it can be set in “n2 Output Mode”.
NOTE