7.2.15 Data Setting Enable / Inhibit (Refer to 6.3.13)
• This function makes it possible to inhibit all data changes (except L1). However,
automatic zero adjustment can be made if the automatic zero-adjustment function is
enabled (set in C1). Also, totalization presetting can be done if the totalization
presetting function is enabled (set in E4).
Display
Description
Selecting "INHIBIT" in parameter number "L01" disables data setting.
Code
Content
ENABLE
INHIBIT
Data setting enabled.
Data setting disabled.
PARAM
L01:TUNING
INHIBIT
L02:KEY
00
L60:SELF CHECK
GOOD
DATA
ESC
DIAG
DATA
ESC
DIAG
Default
7.2.16 Procedure of Selecting Special Application Items (Refer to 6.3.14)
Displayed
Description
It is possible to open up to item N when "55" is entered in parameter
number "L02."
Code
Content
00
55
Accessible up to L parameters
Accessible up to N parameters
PARAM
L01:TUNING
INHIBIT
L02:KEY
55
L60:SELF CHECK
GOOD
DATA
ESC
DIAG
DATA
ESC
DIAG
Default
7.2.17 Rate Limit (Refer to 6.3.15)
Display
Description
PARAM
N03:RATE LIMIT
5 %
N04:DEAD TIME
0 s
N05:POWER SYNCH
YES
DATA
ESC
DIAG
DATA
ESC
DIAG
Set the rate limit value in parameter number "N03." Range of possible
settings : 0 to 10%.
Set the dead time in parameter number "N04."
Range of possible settings : 0 to 15 seconds.
Default : 0
(Rate limit function off)
* The "N" item can be opened by entering "55" in parameter number "L02".
7.2.18 Pulsating Flow (Refer to 6.3.16)
Display
Description
PARAM
N04:DEAD TIME
0
N05:POWER SYNCH
YES
N06:PULSING FLOW
YES
DATA
ESC
DIAG
Select “YES” (pulsating flow counteraction) in parameter number
“N06” (Default : NO)
Code
Content
NO
YES
Normal
Counteracting pulsating flow
* The "N" item can be opened by entering "55" in parameter number "L02".