![YOKOGAWA AE100D Instruction Manual Download Page 203](http://html.mh-extra.com/html/yokogawa/ae100d/ae100d_instruction-manual_911083203.webp)
IM 1F6A0-01E
5-16
5. BASIC OPERATING PROCEDURES
5.6.4 Entering, Setting and Sending Data
The data which are input with the keys are set in the HART
Communicator by pressing ENTER (F4). Then, by pressing
SEND (F2), the data are sent to the YEWFLO. Note that the
data are not set in the YEWFLO if SEND (F2) is not
pressed. All the data set with the HART Communicator is
held in memory unless power is turned off, so every data can
be sent to the YEWFLO at one lot.
Operation
Entering data on the “Tag” setting display.
On alphabetic characters, only capital letters can be used for
setting Tag No. with HART Communicator.
1. Device setup
3. Basic setup
5. Tag
DYF :
Tag
HELP
DEL
ESC
ENTER
1
Call up “Tag” setting display.
On the setting display shown above, enter the data as follows:
Example:
Set “FIC-1A”.
5. Tag
DEF
8
F
F
I
C
-
1
A
GHI
9
F I
ABC
7
F I C
* : +
–
F I C -
STU
1
F I C - 1
ABC
7
F I C - 1 A
Character to
be entered
Operation
Display
F050603_1.EPS
DYF :
Tag
FIC-1A
HELP
DEL
ESC
ENTER
1
Display
Operation
F4
HELP
SEND
HOME
ENTER
2
Press ENTER (F4) to
set the data in the
HART Communicator
after entering the data.
Press SEND (F2) to
send the data to the
YEWFLO.
* is flashing during
communication.
label changed
to label, and
the transmission is
completed.
Press HOME (F3),
and return “Online
Menu”.
HELP
SAVE
HOME
ENTER
3
(ENTER)
F2
(SEND)
SEND
SAVE
DYF :
Basic setup
1 Tag
2 Span
3 PV Damp
4 Device info
DYF :
Basic setup
1 Tag
2 Span
3 PV Damp
4 Device info
F050603_2.EPS
5.6.5 Parameters Configuration
IMPORTANT
Do not turn off the YEWFLO just after HART Commu-
nicator settings (sending) have been made. If the
YEWFLO is turned off less than 30 seconds after
parameters have been set, the set data will not be
stored and the data returns to previous settings.
Parameters of HART Communicator is constructed hierarchi-
cally. The menu tree for Online menu is shown in 5.6.10
Menu Tree.
See appendix “Parameter Summary” about the usage of each
parameter. Note the differences between parameters on
YEWFLO display and those on HART Communicator.
The Online menu summary is shown below.
Table 5.6.1 Online Menu Summary
No.
Display Item
Contents
1
2
3
4
Device setup
PV
PV AO
PV URV
Set parameters for YEWFLO.
Display process value in engineering unit.
Display analog output in mA.
Display set span in engineering unit.
T050601.EPS