5. BASIC OPERATING PROCEDURES
IM 1E7B0-02E
5-1
5.
BASIC OPERATING PROCEDURES
All data settings can be performed with the three keys on the front panel (SET, SHIFT and
INC) or using a handheld BRAIN (BT) terminal.
The following sections describe basic data configurations and how to use the three panel
keys. (See chapter 7 for information on BT operations.)
5.1 Liquid Crystal Display
Figure 5.1 shows the configuration of the ADMAG AE display panel (if equipped).
%
l gal m
3
/h /m
SET
SHIFT
INC
(The figure shows display when fully lit)
(1) LED (red)
Normal operation : off
(2) Delimiter
(3) Unit Display
Units displayed:
l : liter
gal : gallon
m
3
: cubic meters
/h: hours
/m: minutes
If other units are to be used,
attach labels giving the specific unit.
(7) Setting Keys
(6) Decimal Points
(5) Data Display
(4) Unit Label Location
Figure5.1 Configuration of Display
(1) LED (red)
: This LED is off during normal operation and flashes
when an alarm condition has occurred.
(2) Delimiter
: The delimiter " : " (colon) indicates that the displayed data is
in setting mode.
(3) Unit Display
: Displays flow rate units. In order to display other units, the
required unit label should be selected from the provided data
sheets and attached as shown.
(4) Unit Label Location
: To display units not on the LCD, select the required label
from the provided data sheets and attach it here.
(5) Data Display
: Displays flow rate data, setting data and type of alarm
generated.
(6) Decimal Point
: Displays decimal point in the data
(7) Setting Keys
: These keys are used to change flow rate data displays and
type of setting data.