Replacing the cartridge in the optional solvent-selection valve
September 26, 2013, 715004208 Rev. A
99
Replacing the cartridge in the optional solvent-selection
valve
Required materials
•
Gloves: clean, powder-free, chemical-resistant
•
Solvent-selection valve cartridge
Required tool
Allen wrench, 2-mm (startup kit)
To replace the cartridge in the optional solvent-selection valve:
1. In the ACQUITY UPLC Console, select Isocratic Solvent Manager from
the system tree.
2. Ensure the solvent-selection valve is set to solvent 6.
3. Move the solvent bottles to a location below the ISM.
Warning:
To prevent injury, always observe Good Laboratory Practice
when you handle solvents, change tubing, or operate the isocratic
solvent manager. Consult the Material Safety Data Sheets regarding
the solvents you use.
Caution:
To prevent contamination to system components, wear clean,
chemical-resistant, powder-free gloves when replacing the
solvent-selection valve cartridge.
Warning:
To avoid solvent spills, move the solvent bottles to a
location below the ISM.