Installing the optional ACQUITY UPC
2
Flow-splitter kits
September 26, 2013, 715004208 Rev. A
53
8. Reconnect the vacuum tubing that was removed in
to the back of
the degassing chamber that is now attached to the flow-splitter face
plate assembly.
9. Install the Flow-splitter assembly in the front face of the ISM chassis.
Secure the face plate to the ISM with the screws removed in
,
using the T10 TORX driver to tighten the screws.
Location of
vacuum tubing
Screw (5)
Flow-splitter
assembly