Installing the optional Post-column Addition kit
September 26, 2013, 715004208 Rev. A
63
2. Secure the post-column addition bracket with the two bracket screws
that are provided in the Post-Column Addition kit, and then use the T10
TORX driver to tighten them.
3. Connect the tubing as shown in the following diagram.
Notes:
•
Remove the black O-rings prior to installing the tubing. The O-rings
on the tubing are used to hold the fittings in place during shipping.
•
When making the tubing connection to inlet 2, connect the gold
fitting to the ISM filter and the PEEK fitting to inlet 2.
Tip:
To prevent contamination, install fluoropolymer plugs when the
post-column addition kit is not in use.
Post-column
Addition bracket
Blank plate
Bracket screw (2)
Inlet 1:
From optical
detector
Inlet 2:
From ISM
filter
Outlet 1:
To QDa detector
source