11. Disconnect the pressure transducer cable from the bulkhead by
squeezing on the tabs and pulling gently.
12. Using the T27 TORX driver, loosen the two head bolts 1/2-turn. The
bolts are accessible from the front of the pressure transducer.
Pressure transducer
cable connector
Tabs
Head bolt (2)