3 Maintenance Procedures
130
September 26, 2013, 715004208 Rev. A
To remove the accumulator head seals:
1. Stand the head upright on a clean surface.
2. Using the T27 TORX driver, completely loosen the two head bolts to
release the support plate from the pump head.
Requirement:
If you remove the transducer and head bolts, be sure to
reuse the head bolt washers when reassembling the pump head.
3. Lift the pump head from the support plate.
Head bolt (2)
Support plate
Pump head
Seal wash housing