Priming the ISM
September 26, 2013, 715004208 Rev. A
69
2. Follow the stainless-steel solvent vent tubing from port 4 on the vent
valve, and lift it from the drip tray.
3. Push the syringe plunger fully into the syringe barrel.
4. Connect the tubing adapter to the syringe.
5. Connect the syringe assembly to a short length of Pharmed tubing, and
then connect the short length of Pharmed tubing to the solvent vent
tubing you lifted from the drip tray in
.
6. In the ACQUITY Console, select Isocratic Solvent Manager from the
system tree.
7. In the isocratic solvent manager information window, click Control >
Prime solvents.
8. In the Prime Solvents dialog box, select the line you want to prime.
Solvent vent tubing
Vent valve
Drip tray