Installing the optional leak sensor
September 26, 2013, 715004208 Rev. A
33
Installing the optional leak sensor
The ISM can employ an optional leak sensor for detecting and collecting
leakage.
Required materials
•
Gloves: clean, powder-free, chemical-resistant
•
Leak sensor
To install the leak sensor:
1. Power-off the ISM.
2. Open the door of the ISM, gently pulling its right-hand edge toward you.
Warning:
To prevent injury, always observe Good Laboratory Practices
when you handle solvents, change tubing, or operate the ISM. Consult
the Material Safety Data Sheets regarding the solvents you use.
Warning:
The leak sensor can be contaminated with biohazardous
and/or toxic materials. Always wear clean, chemical-resistant,
powder-free gloves when performing this procedure.
Caution:
To avoid damaging electrical parts, never disconnect an
electrical assembly while power is applied to an instrument. To
completely interrupt power, set the power switch to Off, and then
unplug the power cord from the AC source. Wait 10 seconds thereafter
before you disconnect an assembly.