Removing the accumulator head and replacing its seals
September 26, 2013, 715004208 Rev. A
125
8. Remove the seal wash tubing secured to the seal wash housing by
barbed fittings, by using a tool or by pulling on the tubing as close to the
pump head as possible.
9. Using the 1/4-inch open-end wrench, disconnect the outlet tubing from
the transducer.
Seal wash
tubing
Seal wash
tubing
Transducer