Replacing the restrictor module in the optional ACQUITY UPC
2
Flow-splitter kit
September 26, 2013, 715004208 Rev. A
89
Note:
If you are replacing the restrictor module in the ACQUITY UPC
2
Triple Detection Flow-splitter kit, you must also disconnect the outlet 4
tubing from the restrictor module.
2. Using the T10 TORX driver, loosen the two captive screws on the
restrictor module, and then remove the module.
3. Insert the new restrictor module into the flow-splitter assembly.
4. Using the T10 TORX driver, tighten the captive screws on the restrictor
module to the extent possible.
5. Reconnect inlet 3, outlet 2, and outlet 3 to the new restrictor module.
Outlet 2:
To Convergence
Manager tee
Inlet 3
Outlet 3:
To QDa detector
source
Restrictor module
Captive screws