Replacing the low-pressure filter cartridge
September 26, 2013, 715004208 Rev. A
95
9. Insert the low-pressure filter cartridge with tubing into the ferrule
holder fitting.
10. Screw the cap nut onto the ferrule holder fitting, ensuring that the tube
is bottomed out in the in-line filter cartridge. Finger-tighten it to the
extent possible.
11. Return the solvent bottles to their original location.
12. Power-on the isocratic solvent manager.
13. Prime the isocratic solvent manager (see
Ferrule holder fitting
Low-pressure filter