Resolving leak sensor errors
September 26, 2013, 715004208 Rev. A
39
10. Align the leak sensor’s T-bar with the slot in the side of the leak sensor
reservoir, and slide the leak sensor into place.
11. If you detached the connector from the receptacle on the front of the
instrument, reattach it.
12. Power-on the ISM.
13. In the Instrument console, select Isocratic Solvent Manager from the
system tree.
14. In the ISM information window, click Control > Reset ISM, to reset the
ISM.
Slot in leak
sensor reservoir
T-bar
Leak sensor
connector receptacle