AutoSamplers
Varian, Inc.
User Manual 450-GC
Page: 109
14. Remove protective shipping materials from the Injector carriage and install
the syringe into the AutoSampler Injector tower by opening the clear plastic
door, carefully inserting the syringe into its holder and the needle through
its guide, then fasten the plunger screw on the top and snap the retaining
clip over the syringe.
15. Close the clear plastic
sampler door.
16. Replace the left side top cover.
17. Place Waste and Solvent vial (see here for large Solvent / Wash on page
111 vial).
18. Place a capped vial into the Vial 0 location if you have a CP-8400 or the
Vial 1 location if you have a CP-8410.
19. Place the CP-8400 sample-carrousel onto the base or place the CP-8410
sample tray oriented so that Vial 1 is under the arc traced by the tower.
If you re-locate the CP-8400/8410 to the rear position, you
MUST
recalibrate the Injector positions. You will not be automatically
prompted to recalibrate the Injector positions. If you fail to
re-calibrate the Injector positions, the AutoSampler may miss the
Injector, resulti ng in bent syrin ge needles.
Plunger Retaining Nut
(unscrew to remove
syringe and tighten
when installing)
Syringe Retaining Clip
(lift up to remove
syringe)