AutoSamplers
Varian, Inc.
User Manual 450-GC
Page: 119
SYRINGE CHECK/REPLACE
For changing the syringe
size
continue on the next page.
If the syringe has to be replaced by the same size follow the procedure below:
1. Press the SETUP
key and select CALIBRATION, select on the next
page Sampler Change Syringe.
2. The next massage will pop-up and the Sampler Tower will go to his Syringe
Replace/Check Syringe position.
3. Check or Replace the Syringe when the above pop-up message appears.
4. After syringe replacement the syringe plunger stroke counter can be set to
zero, press OK.
5. After Syringe check (no reset of the counter necessary), press Cancel.