Technical Tips
Varian, Inc.
User Manual 450-GC
Page: 459
Sympto m
Cause
Remedy
Check carrier gas flow.
Ghost peaks during temperature
programming, no injection made
Contaminated Injector.
Replace or clean Injector
insert.
Bake out Injector per
Section 4.
Leading peak
Overloaded column.
Carrier gas flow restricted.
Decrease sample size or
increase dilution; use
column with thicker film.
Check column
connections and Injector
insert for obstructions.
Long solvent peak tail
Injector temperature too high.
Injector leakage.
Column temperature too high.
Lower initial Injector
temperature.
Reseat column to glass
insert. See Section 3.
Decrease column
temperature.
No sample peak or peak too
small
Leak at column inlet.
Syringe is leaking.
Broken column.
Injector is leaking.
Check for leaks (Section
4, paragraph 4.2).
Repair or replace syringe.
Replace or repair column.
Check or replace septum
or Injector.
Rising baseline when
temperature programming
Column temperature limit
is exceeded.
Contamination.
Reduce maximum
temperature
Use column with
appropriate coating
thickness for sample. For
relatively non-volatile
compounds that require a
high retention
temperature, choose a
0.1
μ
coating thickness.
Check gas filters. Correct
all leaks, checking for the
following:
(a)make-up gas line union
(b)column-to-detector
connection
(c)detector insert-to-
detector crossover
connection
(d)detector crossover-to-
Summary of Contents for 450-GC
Page 92: ...AutoSamplers Page 152 User Manual 450 GC Varian Inc...
Page 144: ...Detectors Flame Ionization Detector FID Page 294 User Manual 450 GC Varian Inc...
Page 177: ...Maintenance Page 438 User Manual 450 GC Varian Inc...
Page 179: ...Shipping Cleaning and Disposal Instructions Page 440 User Manual 450 GC Varian Inc...
Page 201: ......