AutoSamplers
Page: 114
User Manual 450-GC
Varian, Inc.
1. Place the Calibration tool in Vial 0 or Vial 1
2. Press the SETUP
key and select CALIBRATION, select on the next
page Sampler CALIBRATE.
3. The calibration screen will be visible.
Calibrating the CP-8400 Vial 0 position involves manipulating the positions of 3
different components (Tower, Syringe and Carrousel).
Calibrating the CP-8410 Vial 1 position involves manipulating the positions of 2
different components (Tower and Syringe).
4. Press “Start Calibration”. The CP-8400/8410 will rotate the tower, rotate the
sample-carrousel (not for the CP-8410) and lower the syringe sled to the
approximate location of Vial 0 (or Vial 1 if CP-8410). After the