Installation
Installation
Page:
Page: 14
14
User
User Manual
Manual 450-GC
450-GC
Varian,
Varian, Inc.
Inc.
Nuts and Ferrules for Capillary Columns
Nuts and Ferrules for Capillary Columns
Capillary columns are generally installed with a
Capillary columns are generally installed with a capillary
capillary
column nut (P/N
column nut (P/N 03-94955
03-949551-00) and a
1-00) and a single reversed ferrule,
single reversed ferrule,
as shown in the diagram. For
as shown in the diagram. For connection to the TCD, a
connection to the TCD, a
special capillary adapter (P/N 03-925605-01) is used.
special capillary adapter (P/N 03-925605-01) is used.
A Quick Conn
A Quick Connect Kit (P/N 03-9253
ect Kit (P/N 03-925355-90) facilitate
55-90) facilitates easy
s easy
connection of capillary column to Injector
connection of capillary column to Injector and detector. The
and detector. The
kit comes with instructions and contains: column depth scale,
kit comes with instructions and contains: column depth scale,
knurled split nuts (2) and graphite jacketed ferrules (2
knurled split nuts (2) and graphite jacketed ferrules (2 each of
each of
0.4, 0.5 and 0.8 mm ID, for 250 µm, 320 µm and 530 µm
0.4, 0.5 and 0.8 mm ID, for 250 µm, 320 µm and 530 µm
columns respectively).
columns respectively).
Three types of ferrules are co
Three types of ferrules are common:
mmon: graphite, Vespe
graphite, Vespel® or polyimide and
l® or polyimide and
graphite/Vespel.
graphite/Vespel.
••
Do
Do not
not use
use graphite
graphite with
with ECD
ECD or
or GC/MS
GC/MS systems.
systems.
••
Ferrule
Ferrule ID is
ID is based
based on the
on the diameter
diameter of the
of the capillary
capillary column.
column. See ta
See table
ble
below for 1/16” ferrules.
below for 1/16” ferrules.
Column
Column ID
ID
Holes
Holes
Teflon
Teflon
Max 250°C
Max 250°C
Vespel
Vespel
Max 350°C
Max 350°C
40% Graphite
40% Graphite
/60% Ve
/60% Vespel
spel
Max 400°C
Max 400°C
Graphite
Graphite
Max 450°C
Max 450°C
SilTite
SilTite
Metal, GC/MS
Metal, GC/MS
0.18
0.18 mm
mm ID
ID and
and smaller
smaller
1
1
CR214103
CR214103
CR212103
CR212103
CR213103
CR213103
-
-
--
1
1
CR214104
CR214104
CR212104
CR212104
CR213104
CR213104
CR211104
CR211104
--
0.25 mm ID
0.25 mm ID
2
2
-
-
CR212124
CR212124
CR213124
CR213124
CR211124
CR211124
--
0.25
0.25 mm
mm ID
ID and
and smaller
smaller
1
1
-
-
-
-
-
-
-
-
SG073300
SG073300
1
1
CR214105
CR214105
CR212105
CR213105
CR212105
CR213105
CR211105
CR211105
SG073301
SG073301
0.32 mm ID
0.32 mm ID
2
2
-
-
CR212125
CR212125
CR213125
CR213125
CR211125
CR211125
--
0.53
0.53 mm
mm ID
ID
1
CR214108
1
CR214108
CR212108
CR212108
CR213108
CR213108
CR211108
CR211108
SG073302
SG073302
Table 4: 1/16” Column Ferrules
Table 4: 1/16” Column Ferrules
Reversed
Reversed
Ferrule
Ferrule
Capillary
Capillary
Column
Column
Nut
Nut
Fused Silica
Fused Silica
Column
Column
Summary of Contents for 450-GC
Page 92: ...AutoSamplers Page 152 User Manual 450 GC Varian Inc...
Page 144: ...Detectors Flame Ionization Detector FID Page 294 User Manual 450 GC Varian Inc...
Page 177: ...Maintenance Page 438 User Manual 450 GC Varian Inc...
Page 179: ...Shipping Cleaning and Disposal Instructions Page 440 User Manual 450 GC Varian Inc...
Page 201: ......