AutoSamplers
Page: 116
User Manual 450-GC
Varian, Inc.
GC Injector calibration
1. Press the SETUP
key and select CALIBRATION, select on the next
page Sampler CALIBRATE.
2. The calibration screen will be visible.
3. Before starting the calibration identify Injector 1 and Injector 2.
Injector 1 is the first Injector the Tower will encounter when it rotates
clockwise to reach the Injector(s). The second Injector is Injector 2. Please
notice the difference when the sampler is mounted in the Front or Rear
(next page) position.
Injector 2
Injector 1
Sampler mounted in
the Front position