Injectors – 1177 Split/Splitless Capillary Injector
Page: 164
User Manual 450-GC
Varian, Inc.
Gas Saver
For type EFC21 and EFC25 a specific Gas Saver Flow and a timeout period
may be set. Press the setup
key and click on the Injector tab now choose
the correct Injector (Page 1 or Page 2 or Page 3).
A time entry of 0 turns the Gas Saver feature off
. If a non-zero time is
entered, the Gas Saver timer will start incrementing. If a run is not started or a
new method is not loaded or download by Workstation, the set Gas Saver flow
rate will be activated (the instrument will go into stand-by). A message will be
logged in the instrument log.
Stand-by is canceled by loading a method or
pressing the end stand-by key.