Standard On-Column and Standard Split/Splitless Sampling Modes
Internal Standard i s no t permitt ed with Standard Sampling modes.
Standard mode parameters are set for samples dissolved in common solvents
such as iso-octane.
The Standard sampling mode for
On-Column
performs the following
sequence of actions:
Sampler
sequence
Standard On-column - mode
SyringeSize
(µL)
1
2
5
10
50
100
250
1
If used: Pre-Injection Solvent Clean Flushes
(solvent
→
waste)
- Volume (µL)
0.75
1.5
3.75
7.5
37.5
75
187.5
- Plunger Speed Up (µL/s)
0.5
1
2.5
5
10
20
50
- Plunger Speed Down (µL/s)
5
10
25
50
25
50
62.5
2
If used: Any Pre-Injection Sample Clean
Flushes
(sample
→
waste)
- Volume (µL)
0.75
1.5
3.75
7.5
37.5
75
187.5
- Plunger Speed Up (µL/s)
0.5
1
2.5
5
10
20
50
- Plunger Speed Down (µL/s)
5
10
25
50
25
50
62.5
3
Clean Strokes
(sample
→
sample)
- Plunger Speed Up (µL/s)
0.2
0.4
1
2
10
20
50
- Plunger Speed Down (µL/s)
5
10
25
50
25
50
62.5
- Volume For Fill Strokes (µL)
0.75
1.5
3.75
7.5
37.5
75
187.5
- Number Of Clean Strokes
5
5
5
5
5
5
5
- Viscosity Delay (s)
0.0
0.0
0.0
0.0
0.0
0.0
0.0
4
Draw Sample
- Plunger Speed During Fill (µL/s)
0.2
0.4
1
2
10
20
50
- Viscosity Delay (s)
0.0
0.0
0.0
0.0
0.0
0.0
0.0
Al l Cl eanStrokes - and Fl ush-Volumes are limit ed to 75% of the sy ringe
capacity.
Summary of Contents for 450-GC
Page 92: ...AutoSamplers Page 152 User Manual 450 GC Varian Inc...
Page 144: ...Detectors Flame Ionization Detector FID Page 294 User Manual 450 GC Varian Inc...
Page 177: ...Maintenance Page 438 User Manual 450 GC Varian Inc...
Page 179: ...Shipping Cleaning and Disposal Instructions Page 440 User Manual 450 GC Varian Inc...
Page 201: ......