Appendix A: Error Codes
Page: 488
User Manual 450-GC
Varian, Inc.
416
Sampler Tray is missing
Alert
Only on 8410 sampler
Notification Message?
Place tray in position
418
Sampler Vial is unreachable
Alert
Only on 8410 sampler, tray has
been turned vial becomes
unreachable for the tower
Notification Message
Calibrate AS, place vial,
solve error
423
Sampler Vial sensor bad
Alert
Defective sensor
Electronics off (all)
H2 flow set to 0.0
mL/min
Replace vial sensor
424
Sampler Column standby timeout expired
Event
Standby timeout expired
Activate method, reset
time out
Activate method, reset time
out
426
Sampler Plunger stuck on sensor
Alert
(errors not reproducible)
Electronics off (all)
H2 flow set to 0.0
mL/min
-
427
Sampler Syringe stuck on sensor
Alert
(errors not reproducible)
Electronics off (all)
H2 flow set to 0.0
mL/min
-
428
Sampler Tower stuck on sensor
Alert
(errors not reproducible)
Electronics off (all)
H2 flow set to 0.0
mL/min
-
429
Sampler Carrousel stuck on sensor
Alert
(errors not reproducible)
Electronics off (all)
H2 flow set to 0.0
mL/min
-
434
Sampler EDS Checksum is bad
Alert
-
-
-
436
IOP Communication failure
Error
IOP Communication failure
Notification Message?
Replace/fix IOP
437
Battery voltage out of range
Event
Battery voltage < 2V or > 4V
error
-
-
439
PGA Calib Failed
Error
IOPSH (measuring system)
-
Recalibrate PGA
440
Temp Calib Failed
Error
-
-
Recalibrate temp
442
Press out of tolerance
Error
Ambient pressure < 15kPa or >
115pKa
error
-
-
443
Boardtemp out of tolerance
Error
Currently in use, only logged
-
GC main board
Out of Tolerance (red_id)
444
P10V ref out of tolerance
Error
-
445
AIN10V out of tolerance
Error
-
446
P5V ref out of tolerance
Error
-
447
P2V5 ref out of tolerance
Error
444 to 465 Refvolt out of Tolerance
(all voltages on mainboard, part of
measuring system)
Hazardous by default, basis of all
measurement in system
-
Check power supply
requirements/ replace main
board electronics
Summary of Contents for 450-GC
Page 92: ...AutoSamplers Page 152 User Manual 450 GC Varian Inc...
Page 144: ...Detectors Flame Ionization Detector FID Page 294 User Manual 450 GC Varian Inc...
Page 177: ...Maintenance Page 438 User Manual 450 GC Varian Inc...
Page 179: ...Shipping Cleaning and Disposal Instructions Page 440 User Manual 450 GC Varian Inc...
Page 201: ......