Installation
Installation
Varian,
Varian, Inc.
Inc.
User
User Manual
Manual 450-GC
450-GC
Page:
Page: 17
17
13.
13. Gently push the column into the Injector until the mark on the column is
Gently push the column into the Injector until the mark on the column is
aligned with the bottom edge of the column nut.
aligned with the bottom edge of the column nut.
14.
14. Tighten the column nut by hand until it is finger-tight. Hold the column and
Tighten the column nut by hand until it is finger-tight. Hold the column and
use a 5/16" wrench to tighten the column nut in place. Tighten the column
use a 5/16" wrench to tighten the column nut in place. Tighten the column
nut only enough to seal the column and hold it firmly in place.
nut only enough to seal the column and hold it firmly in place.
An optional Capillary Column Quick Connect Kit
An optional Capillary Column Quick Connect Kit is available that
is available that
simplifies the installation of a capillary column into Injectors and
simplifies the installation of a capillary column into Injectors and
detectors. It contains split capillary column nuts, re-usable graphite
detectors. It contains split capillary column nuts, re-usable graphite
ferrules
ferrules (for
(for 250
250 m,
m, 320
320 m,
m, 530
530 m
m ID
ID columns)
columns) and
and a
a column
column
depth scale. The ferrule is set on the column using this depth scale.
depth scale. The ferrule is set on the column using this depth scale.
No felt pen or typewriter correction fluid is needed (The part number
No felt pen or typewriter correction fluid is needed (The part number
of the Quick Connect Kit is 03-925751-90).
of the Quick Connect Kit is 03-925751-90).
Tips for Tightening Column Ferrules
Tips for Tightening Column Ferrules
Refer to the
Refer to the Table 6
Table 6
below
below
for instructions to
for instructions to tighten the capillary column nut
tighten the capillary column nut
with a variety of
with a variety of ferrule types. Do not over
ferrule types. Do not over tighten capillary column nuts use a
tighten capillary column nuts use a
new ferrule if the column does not
new ferrule if the column does not seal. Over tightening capillary nuts can
seal. Over tightening capillary nuts can
destroy some ferrules,
destroy some ferrules, particularly graphite.
particularly graphite.
Fe
Ferrul
rrul e
e Type
Type
Recommendation
Recommendation
Comments
Comments
New
New
graphite
graphite ferrule
ferrule
¼
¼ -
- turn
turn past
past finger-tight
finger-tight
¼ - to ½ -turn
¼ - to ½ -turn past finger-tight may be
past finger-tight may be
required to achieve a seal.
required to achieve a seal.
Used
Used
graphite
graphite ferrule
ferrule
¼
¼ -
- turn
turn past
past finger-tight
finger-tight
If a different size of column is used,
If a different size of column is used,
more tightening may be required.
more tightening may be required.
New
New
graphite-Vespel®
graphite-Vespel®
ferrule
ferrule
¼ - to ½ -turn past finger-tight
¼ - to ½ -turn past finger-tight
May require retightening after first or
May require retightening after first or
second programmed run. Over
second programmed run. Over
tightening will destroy ferrule and seal.
tightening will destroy ferrule and seal.
Used
Used
graphite-Vespel®
graphite-Vespel®
ferrule
ferrule
¼
¼ -
- turn
turn past
past finger-tight
finger-tight
Re-use
Re-use only
only on
on same
same size
size column.
column.
1177/1079 PTV: ¼ - to
1177/1079 PTV: ¼ - to ½ -turn past
½ -turn past
finger-tight.
finger-tight.
May require retightening after first or
May require retightening after first or
second programmed run.
second programmed run.
New
New
Vespel®
Vespel® ferrule
ferrule
Not recommended for 1041/1061/1093
Not recommended for 1041/1061/1093
1177/1079 PTV: ¼ -turn past finger-
1177/1079 PTV: ¼ -turn past finger-
tight.
tight.
Reuse only on same size column
Reuse only on same size column
Used
Used
Vespel®
Vespel® ferrule
ferrule
Not recommended for 1041/1061/1093
Not recommended for 1041/1061/1093
Table 6: General Tightening and Re-tightening for Common Ferrules
Table 6: General Tightening and Re-tightening for Common Ferrules
Summary of Contents for 450-GC
Page 92: ...AutoSamplers Page 152 User Manual 450 GC Varian Inc...
Page 144: ...Detectors Flame Ionization Detector FID Page 294 User Manual 450 GC Varian Inc...
Page 177: ...Maintenance Page 438 User Manual 450 GC Varian Inc...
Page 179: ...Shipping Cleaning and Disposal Instructions Page 440 User Manual 450 GC Varian Inc...
Page 201: ......