9. Insert the signal probe into the upper arm of the detector tower. The probe
clip should fit around the tapered section on the collector tube tightly
enough to exert a downward force (see Figure 47). Secure the signal probe
by tightening the knurled nut. Make sure the notch in the detector tower arm
does not cut the O-ring seal.
10. Connect the igniter cable to the igniter probe and the signal cable to the
signal probe.
Figure 46: Orientation of Igniter Probe Elements in FID Tower
Figure 47: FID Cross-sectional View
Collector Tube
(03-949587-00)
Signal Probe
Notch
Ignitor Probe
Insulator
Detector Tower
Spring Clip
Ignitor Coil
TOP VIEW
Flame Tip
Ignitor Coil
Spring Clip
Flame Tip
Assembly