AutoSamplers
Varian, Inc.
User Manual 450-GC
Page: 107
8. Route the CP-8400/8410 power/communication cable through the Sample
top-cover gap.
9. Install the Injector locator plate by gently pushing the
circular holes over the Injector(s). Do not use
excessive force. If the Injector(s) do not fit
conveniently, loosen the screws holding the
Injector(s) to the 450-GC chassis and adjust the
Injector(s) positions so that they conveniently fit into
the locator plate. The top cover has small access
holes to allow you to loosen the retaining screws on
the Injectors. You may need to unscrew the Injector
nut and remove the inject switch to have unimpeded
access to the Injector retaining screws; replace the
Injector nut (if removed) so that you have proper
alignment when the locator plate is installed. Once
the Injector(s) are properly aligned, screw down the
alignment plate and then screw down the Injector
retaining screws. Remove the locator plate, replace
the inject switch on each of the Injector(s) and
reinstall the locator plate.
10. The CP-8400/8410 is aligned to the Top-Cover of the
450-GC by a 2 pins protruding from the bottom of the
AutoSampler/AutoInjector base. These 2 pins will fit into holes on the Top-
Cover.
Injector locator plate