Shipping, Cleaning and Disposal Instructions
Varian, Inc.
User Manual 450-GC
Page: 439
SHIPPING, CLEANING AND DISPOSAL INSTRUCTIONS
SHIPPING INSTRUCTIONS
If you r 450-GC for any reason m ust s end back to the nearest Varian, Inc.
subsidi ary or Varian, Inc. representative, it is v ery important to follow these
additional preparation inst ructions :
1. Place the shipping restrain pin back.
2. Put protection plugs on the sample and/or carrier gas inlets.
3. Remove columns.
CLEANING INSTRUCTIONS
To keep the 450-GC surface clean, refer to the remarks given below:
•
Switch the 450-GC off.
•
Remove the power cable.
•
Put protection plugs on the sample and carrier gas inlets.
•
Put protection plugs on the Column vents.
•
Use a soft (no hard or abrasive) brush to carefully brush away all dust and
dirt.
•
If the outer case is dirty (never clean the inside!) clean it with a soft, clean
cloth dampened with mild detergent.
•
Never use alcohol or thinners to clean the 450-GC, these chemicals can
damage the case.
•
Be careful not to get water on the electronic components.
•
Do not use compressed air to clean.
DISPOSAL INSTRUCTIONS
When the lifetime of the 450-GC or parts of it has reached the end of its useful
life, disposal must be carried out in accordance with all (environment)
regulations applicable in your country.
Summary of Contents for 450-GC
Page 92: ...AutoSamplers Page 152 User Manual 450 GC Varian Inc...
Page 144: ...Detectors Flame Ionization Detector FID Page 294 User Manual 450 GC Varian Inc...
Page 177: ...Maintenance Page 438 User Manual 450 GC Varian Inc...
Page 179: ...Shipping Cleaning and Disposal Instructions Page 440 User Manual 450 GC Varian Inc...
Page 201: ......