Standards and IEC Classifications
T
e
chnical S
pecifications
Force EZ-C Series Service Manual
4-9
Liquid Spillage (IEC 60601-2-2 Clause 44.3)
The Force EZ-C Series Electrosurgical Generator enclosure is constructed so that
liquid spillage in normal use does not wet electrical insulation or other
components which, when wetted, are likely to adversely affect the safety of the
equipment.
Electromagnetic Interference
The Force EZ-C Series Electrosurgical Generator minimizes electromagnetic
interference to video equipment used in the operating room. The generator
complies with the requirements of IEC 61000-4-2.
Electromagnetic Compatibility (IEC 60601-1-2 and
IEC 60601-2-2)
The Force EZ-C Series Electrosurgical Generator meets IEC 60601-1-2 and IEC
60601-2-2 requirements for electromagnetic compatibility.
Voltage Transients (Emergency Generator Mains
Transfer)
The Force EZ-C Series Electrosurgical Generator operates in a safe manner when
you transfer between line AC and an emergency generator voltage source.
Notice
The Force EZ should not be used adjacent to or stacked with equipment other
than specified in the Force EZ User Guide and Service Manual. If adjacent or
stacked use is necessary, the Force EZ should be observed to verify normal
operation in the configuration in which it will be used.
The Force EZ intentionally applies RF energy for diagnosis or treatment during
activation. Observe other electronic medical equipment in the vicinity during the
Force EZ activation for any possible adverse electromagnetic effects. Ensure
adequate separation of electronic medical equipment based on observed
reactions.
The use of accessories, other than specified in the Force EZ User Guide and
Service Manual, may result in increased emissions or decreased immunity of the
Force EZ.