Returning Circuit Boards
Rep
a
ir
P
o
licy and Pr
oc
edur
es
Force EZ-C Series Service Manual
9-3
B.
Thoroughly wipe all surfaces of the generator and power cord with a mild
cleaning solution or disinfectant and a damp cloth. Follow the procedures
approved by your institution or use a validated infection control procedure.
Do not allow fluids to enter the chassis. You cannot sterilize the generator.
Step 3 – Ship the Generator
A.
Attach a tag to the generator that includes the Return Authorization Number
and the information (hospital, phone number, etc.) listed in
Step 1 – Obtain a
Return Authorization Number
.
B.
Be sure the generator is completely dry before you pack it for shipment. Pack
it in its original shipping container, if available.
C.
Ship the generator, prepaid, to the Valleylab Service Center.
Returning Circuit Boards
Pack circuit boards for shipment as follows:
1.
Place each circuit board in an electrostatic discharge (ESD) bag or container.
2.
Provide a separate packing container for each circuit board.
3.
Attach a tag to the container that includes the Return Authorization Number
and the information (hospital, phone number, etc.) listed in
Step 1-Obtain a
Return Authorization Number
.
4.
Ship the circuit board prepaid to the Valleylab Service Center.