Responding to System Alarms
7-14
Force EZ-C Series Service Manual
90
95
Generator model number in master
microcontroller Flash Memory not
Force EZ-C Series Electrosurgical
Generator model number.
Generator model number in feedback
microcontroller Flash Memory not
Force EZ-C Series Electrosurgical
Generator model number.
Turn the power switch off (O) then on (I) again.
If error reappears, replace the control board.
Refer to
Control Board Replacement
in Section 8.
100
101
102
103
104
105
109
110
111
112
113
114
115
116
Software malfunction.
rem_update_state
value outside state range.
Software malfunction.
rem_pad_state
value outside state range.
Software malfunction.
rem_state
value in
re_rem_control
( ) outside state range.
Software malfunction.
rem_state
value in
re_non_rem_control
( ) outside state
range.
Software malfunction.
rem_flash_state
value outside state range.
Software malfunction.
rem_led_state
value outside state range.
Software malfunction.
coag_state
value in
ca_econ_cal
() outside state range.
Software malfunction.
cal_state
value in
ca_generator_setup
( ) outside state
range.
Software malfunction.
cal_state
value in
ca_clock_date_setup
( ) outside state
range.
Software malfunction.
cal_mode
value in
ca_clock_date_setup
( ) outside state
range.
Software malfunction.
cal_state
value in
ca_clock_time_setup
( ) outside state
range.
Software malfunction.
cal_mode
value in
ca_clock_time_setup
( ) outside state
range
Software malfunction.
cal_state
value in
ca_coag_leakage_cal
( ) outside state
range.
Software malfunction.
cal_state
value in
ca_econ_cal
( ) outside state range.
Turn off, then turn on the generator. If the alarm
number reappears, record the number and call the
Valleylab Service Center.
Number
Description
Recommended Action